Understanding the Gorlin formula calculator — concept and principle
The Gorlin formula calculates the area of a narrowed (stenotic) aortic or mitral valve from measurements taken during cardiac catheterisation: cardiac output, heart rate, the time the valve is open in each beat and the mean pressure gradient across it. Cardiologists use it to grade valve stenosis when echocardiography is unclear.
How it is calculated
Flow across the valve (mL/s) = cardiac output (mL/min) ÷ (heart rate × period per beat)
Valve area (cm²) = flow ÷ (C × √mean gradient)
- The period is the systolic ejection period (SEP) for the aortic valve or the diastolic filling period (DFP) for the mitral valve, in seconds per beat.
- C is an empirical constant: 44.3 for the aortic valve and 37.7 (44.3 × 0.85) for the mitral valve.
- The Hakki shortcut, area ≈ cardiac output (L/min) ÷ √gradient, is shown for comparison.
The principle
The formula comes from hydraulics. Flow through an opening equals area times velocity, and by Torricelli's law the velocity of a jet rises with the square root of the pressure drop. Rearranging gives area = flow ÷ (constant × √pressure). Blood only crosses the valve while it is open, so cardiac output is divided by the total opening time per minute to get the true flow rate during that time. The constant absorbs unit conversions, gravity and the narrowing of the jet beyond the opening.
Worked example
Default inputs: aortic valve, cardiac output 4.5 L/min, heart rate 75 bpm, SEP 0.30 s, mean gradient 50 mmHg.
- Flow = 4,500 ÷ (75 × 0.30) = 4,500 ÷ 22.5 = 200 mL/s.
- √50 = 7.071; 44.3 × 7.071 = 313.2.
- Valve area = 200 ÷ 313.2 = 0.639 → 0.64 cm² → Severe (under 1.0 cm²).
- Hakki = 4.5 ÷ 7.071 = 0.64 cm².
| Heart rate (bpm) | Valve area (cm²) |
|---|---|
| 18.8 | 2.55 |
| 37.5 | 1.28 |
| 56.2 | 0.85 |
| 75 | 0.64 |
| 93.8 | 0.51 |
| 112 | 0.43 |
| 150 | 0.32 |
How to read the result
Common grading: over 1.5 cm² mild, 1.0–1.5 cm² moderate, under 1.0 cm² severe. For aortic stenosis many guidelines also use an indexed area and the gradient. A small area with a low gradient and low output needs further study, as the valve may open more with higher flow.
Common mistakes
- Using the aortic constant for the mitral valve.
- Using forward cardiac output when there is also regurgitation, which understates true valve flow.
- Entering the period per minute instead of per beat.
Frequently asked questions
Why is the Gorlin formula less accurate at low cardiac output?
The empirical constant was derived at normal flows; at low flow the formula tends to underestimate the area.
When does the Hakki formula fail?
At fast or slow heart rates, when the opening period per minute differs from the usual value.
Is echo or catheter valve area preferred?
Echo is the first test; catheter measurement is used when echo results conflict with the clinical picture.
Sources: Gorlin R, Gorlin SG, "Hydraulic formula for calculation of the area of the stenotic mitral valve, other cardiac valves, and central circulatory shunts", Am Heart J 1951;41:1–29 · Hakki AH et al., "A simplified valve formula for the calculation of stenotic cardiac valve areas", Circulation 1981;63:1050–1055 · Baumgartner H et al., EAE/ASE recommendations on valve stenosis, Eur J Echocardiogr 2009;10:1–25