Understanding the BOD₅ calculator — concept and principle
Biochemical oxygen demand (BOD₅) is the amount of dissolved oxygen that microorganisms consume while breaking down organic matter in a water sample over five days at 20 °C. Because a strong sample would use up all the oxygen in the bottle, it is diluted with oxygen-saturated dilution water first. This calculator converts the measured oxygen drop in the diluted bottle back to the BOD of the original, undiluted sample.
How it is calculated
BOD₅ (mg/L) = (D₁ − D₂) ÷ P
- D₁: dissolved oxygen of the diluted sample right after preparation (mg/L).
- D₂: dissolved oxygen after 5 days of incubation (mg/L).
- P: decimal fraction of sample in the bottle, sample volume ÷ bottle volume (for example 6 mL in a 300 mL bottle gives 0.02).
This is the unseeded form. If seed microorganisms are added, the seed's own demand is subtracted: BOD₅ = [(D₁ − D₂) − (B₁ − B₂) × f] ÷ P, where B₁ and B₂ are the seed control readings and f is the ratio of seed in the sample bottle to seed in the control.
The principle
Bacteria oxidise biodegradable organic matter and use oxygen in proportion to how much they break down. In a sealed, dark bottle no oxygen enters and none is produced by photosynthesis, so the fall in DO measures microbial consumption alone. Only the fraction P of the bottle is real sample; the rest is dilution water that ideally consumes nothing. Dividing by P scales the oxygen used in the bottle up to what the full-strength sample would demand.
Worked example
- Inputs: D₁ = 8.5 mg/L, D₂ = 4.0 mg/L, P = 0.02.
- Oxygen depletion: 8.5 − 4.0 = 4.5 mg/L.
- BOD₅ = 4.5 ÷ 0.02 = 225 mg/L.
- Validity check: the depletion (4.5 mg/L) is at least 2 mg/L and the residual DO (4.0 mg/L) is at least 1 mg/L, so the dilution is acceptable under Standard Methods criteria.
| Sample fraction P (sample vol ÷ bottle vol) | BOD₅ (mg/L) |
|---|---|
| 0.005 | 900 |
| 0.01 | 450 |
| 0.015 | 300 |
| 0.02 | 225 |
| 0.025 | 180 |
| 0.03 | 150 |
| 0.04 | 112.5 |
How to read the result
Untreated domestic wastewater typically has a BOD₅ of a few hundred mg/L, so 225 mg/L is a normal raw-sewage value. Secondary treatment brings it down to the low tens: the EU Urban Waste Water Treatment Directive sets 25 mg/L, and the US secondary treatment standard 30 mg/L as a 30-day average.
Common mistakes
- Entering P as a percentage (2) instead of a fraction (0.02).
- Accepting bottles where depletion is below 2 mg/L or residual DO below 1 mg/L.
- Forgetting the seed correction when seed was added.
Frequently asked questions
Why five days?
It became a convention in early 20th-century Britain, where five days was long enough for substantial oxidation and considered longer than river water took to reach the sea.
What is the difference between BOD and COD?
COD uses a strong chemical oxidant and oxidises most organic matter in a few hours, so COD is usually higher than BOD.
How do I choose P?
Prepare several dilutions so that at least one depletes 2 mg/L and leaves 1 mg/L.
Sources: APHA, AWWA and WEF, Standard Methods for the Examination of Water and Wastewater, Method 5210 B · Council Directive 91/271/EEC concerning urban waste-water treatment, Annex I · US EPA, 40 CFR Part 133, Secondary Treatment Regulation